Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Beckman HD160: (1 year) CAL VER /5520 DATE: 02-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 34 NUMBER OF LINES: 136 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P J F W 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: NONE 1.004 DISP Temperature: -20 - 50degC 1.004 DISP Humidity: <80% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {AC VOLTAGE} 1.007 DISP Set the UUT to VAC. 1.007 DISP 1.007 DISP *NOTE* If the "battery" symbol appears in the LCD 1.007 DISP display, replace the battery before continuing 1.007 DISP Connect the 5520A to the UUT as follows: 1.007 DISP [32] 5520A NORMAL HI to UUT V-Ohm 1.007 DISP [32] 5520A NORMAL LO to UUT COM 1.008 HEAD AC VOLTAGE: {10V Range} 1.009 5520 10 0.000V -0U +0.005U 2W 2.001 5520 10 3.500V 0.058U 10kH SI 2W 3.001 5520 10 4.100V 0.063U 60H SI 2W 4.001 5520 10 4.100V 0.063U 10kH SI 2W 5.001 HEAD AC VOLTAGE: {100V Range} 5.002 5520 100 35.00V 0.54U 60H SI 2W 6.001 5520 100 35.00V 0.54U 1kH SI 2W 7.001 5520 100 41.00V 0.63U 60H SI 2W 8.001 5520 100 41.00V 0.63U 1kH SI 2W 9.001 HEAD AC VOLTAGE: {1000V Range} 9.002 5520 1000 350.0V 5.3U 60H SI 2W 10.001 5520 1000 350.0V 5.3U 1kH SI 2W 11.001 5520 1000 900.0V 13.6U 60H SI 2W 12.001 5520 1000 900.0V 13.6U 1kH SI 2W 13.001 HEAD {DC VOLTAGE} 13.002 DISP Set the UUT to VDC. 13.003 HEAD DC VOLTAGE: {1000mV Range} 13.004 5520 1000 900.0mV 2.6U 2W 14.001 5520 1000 -900.0mV 2.6U 2W 15.001 HEAD DC VOLTAGE: {10V Range} 15.002 5520 10 2.500V 0.007U 2W 16.001 5520 10 9.000V 0.024U 2W 17.001 HEAD DC VOLTAGE: {100V Range} 17.002 5520 100 35.00V 0.10U 2W 18.001 5520 100 42.00V 0.12U 2W 19.001 HEAD DC VOLTAGE: {1000V Range} 19.002 5520 1000 350.0V 1U 2W 20.001 5520 1000 900.0V 2.4U 2W 21.001 HEAD {RESISTANCE} 21.002 DISP Set the UUT to OHMS. 21.003 DISP Connect the 5520A to the UUT as follows: 21.003 DISP [32] 5520A AUX HI to UUT V-Ohm 21.003 DISP [32] 5520A AUX LO to UUT COM 21.003 DISP [32] 5520A NORMAL HI to UUT V-Ohm 21.003 DISP [32] 5520A NORMAL LO to UUT COM 21.003 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 21.003 DISP the NORMAL leads to the SENSE leads AT the UUT. 21.004 HEAD RESISTANCE: {1000 Ohm Range} 21.005 5520 1000 900.0Z 6U CW 22.001 HEAD RESISTANCE: {10 kOhm Range} 22.002 5520 10 10.00kZ 0.06U CW 23.001 HEAD RESISTANCE: {100 kOhm Range} 23.002 5520 100 100.0kZ 0.6U CW 24.001 HEAD RESISTANCE: {1000 kOhm Range} 24.002 DISP Connect the 5520A to the UUT as follows: 24.002 DISP [32] 5520A NORMAL HI to UUT V-Ohm 24.002 DISP [32] 5520A NORMAL LO to UUT COM 24.003 5520 1000 900.0kZ 6U 2W 25.001 HEAD RESISTANCE: {40 MOhm Range} 25.002 5520 40 10.00MZ 0.17U 2W 26.001 HEAD {DC LOW CURRENT} 26.002 DISP Set the UUT to A(DC). 26.003 DISP Connect the 5520A to the UUT as follows: 26.003 DISP [32] 5520A AUX HI to UUT 40mA 26.003 DISP [32] 5520A AUX LO to UUT COM 26.004 HEAD DC LOW CURRENT: {10mA Range} 26.005 5520 10 0.900mA 0.012U 2W 27.001 5520 10 9.000mA 0.073U 2W 28.001 HEAD DC LOW CURRENT: {40mA Range} 28.002 5520 40 25.00mA 0.24U 2W 29.001 HEAD {AC LOW CURRENT} 29.002 DISP Set the UUT to A(AC). 29.003 HEAD AC LOW CURRENT: {10mA Range} 29.004 5520 10 9.000mA 0.163U 1kH SI 2W 30.001 HEAD {AC HIGH CURRENT} 30.002 HEAD AC HIGH CURRENT: {10A Range} 30.003 DISP Connect the 5520A to the UUT as follows: 30.003 DISP [32] 5520A AUX HI to UUT 20A 30.003 DISP [32] 5520A AUX LO to UUT COM 30.004 5520 10 1.900A 0.038U 1kH SI 2W 31.001 HEAD AC HIGH CURRENT: {20A Range} 31.002 DISP Connect the 5520A to the UUT as follows: 31.002 DISP [32] 5520A 20A to UUT 20A 31.002 DISP [32] 5520A AUX LO to UUT COM 31.003 5520 20 11.00A 0.22U 1kH SI HC 2W 32.001 HEAD {DC HIGH CURRENT} 32.002 DISP Set the UUT to A(DC) 32.003 HEAD DC HIGH CURRENT: {10A Range} 32.004 DISP Connect the 5520A to the UUT as follows: 32.004 DISP [32] 5520A AUX HI to UUT 20A 32.004 DISP [32] 5520A AUX LO to UUT COM 32.005 5520 10 1.900A 0.019U 2W 33.001 HEAD DC HIGH CURRENT: {20A Range} 33.002 DISP Connect the 5520A to the UUT as follows: 33.002 DISP [32] 5520A 20A to UUT 20A 33.002 DISP [32] 5520A AUX LO to UUT COM 33.003 5520 20 11.0A 0.11U HC 2W 34.001 END